Teilenummer
Entmilitarisierung
Lagerfähigkeit
UOM
NIIN
Informationen zum Teil
Model ht high speed centrifuge
NSN
Nationale Bestandsnummer:
TXT
Beschreibung:
Laboratory Centrifuge
MRC:
An overall measurement between specified points of an item,in distinction from height.ABFY
Overall Depth:
17.000 inches
MRC:
The distance measured in a straight line from the bottom to the top of an item.ABKW
Overall Height:
21.500 inches
MRC:
An overall measurement taken at right angles to the length of an item,in distinction from thickness.ABMK
Overall Width:
14.000 inches
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
The number of complete cyclic changes,per unit of time,for which an item is rated.ACZB
Frequency Rating:
60.0 hertz
MRC:
The name and number of parts supplied with the item which may be required for application.AKYD
Accessory Components And Quantity:
Rotor; 3 wire cord/plug
MRC:
The means used to operate the item.APHE
Operation Method:
Electric
MRC:
Indicates the type of separation unit provided.ATPH
Separation Unit Type:
Flat capillary tube
MRC:
The number of slots in the item.ATPP
Slot Quantity:
8
MRC:
An indication of whether or not a safety cap is included.ATPQ
Safety Cap:
Not included
MRC:
The maximum speed at which the item is designed to operate,expressed in revolutions per minute.ATPR
Maximum Speed Rating In Rpm:
17000.0
MRC:
An indication of whether or not a guard bowl is included.ATPT
Guard Bowl:
Included
MRC:
The total electrical voltage.ELEC
Voltage In Volts:
115.0
MRC:
The number of alternating current phases.FAAZ
Phase:
Single
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
High speed bench type centrifuge; spins 8 50 ml tubes (400 ml total) at speeds up to 17, 000 rpm; detachable control panel for remote operation of the centrifuge-requires iec no. 1857 extension calbe; 8 place 33 degree angle rotor holds round bottom tubes of 28.6 by 107 mm diameter and length; control panel includes stepless variable speed control; combination ammeter/tachometer with switch to select observation of motor speed or current, automatic 2 to 120 minute timer, power switch for timed or continuous operation and pilot light; equipped with hinged steel plate cover and lever latch; guard bowl made of 2 thicknesses of heavy gauge steel; case made of enameled steel; 1/2 hp motor; air cooled
MRC:
Characteristics or qualities of an item,not covered in any other requirement,which are considered essential information for one or more functions excluding nsn assignment.SUPP
Supplementary Features:
Model ht centrifuge from iec part number 3401